자동 압력 교정/조절기 유압용(고정밀도) PPCH

- 유압으로 최대 200 MPa (30,000 psi) - 초정밀 압력 조절 - 자동압력범위 생성 - 0.01% of reading 불확도   Hydraulic pressure controller to 200 MPa (30,000 psi) Hydraulic pressure controller to 200 MPa (30,000 psi) High precision control AutoRange feature For liquid pressures up to 30,000 psi (200 MPa) One or two Q-RPTs in […]

상세 보기

- 유압으로 최대 200 MPa (30,000 psi)
- 초정밀 압력 조절
- 자동압력범위 생성
- 0.01% of reading 불확도

 

Hydraulic pressure controller to 200 MPa (30,000 psi)

  • Hydraulic pressure controller to 200 MPa (30,000 psi)
  • High precision control
  • AutoRange feature
  • For liquid pressures up to 30,000 psi (200 MPa)
  • One or two Q-RPTs in each controller
  • Supports external reference pressure transducers
  • FLASH memory for simple and free embedded software upgrades via Fluke Calibration web-site

PPCH™, is a pressure generator/controller/calibrator for liquid pressure operation from 1 to 200 MPa (150 to 30 000 psi). As with the rest of Fluke Calibration's, PPC family of pressure controllers, the emphasis is on high end performance, minimizing measurement uncertainty and maintaining precise control over a very wide pressure range… in a compact and rugged instrument.

New, individually characterized, quartz reference pressure transducer (Q-RPT) modules increase precision and reduce measurement uncertainty. The AutoRange™ feature supports infinite ranging, automatically optimizing all aspects of operation for the exact range of the DUT being calibrated and taking pressure controller rangeability to a new level.

A unique pressure generation and control system provides unlimited, on-demand pressure, very high control resolution and 10:1 pressure control turndown. Four different control modes are included for maximum versatility.

Open architecture allows reference pressure measurement to be internal to, or remote from, the controller. If desired, the reference can be located at the test measurement point and independently removed for recalibration. With all of this, PPCH opens new doors in automated high pressure hydraulic calibration and test applications.

 

Description
Quartz reference pressure transducer (Q-RPT) modules

PPCH's outstanding pressure measurement specifications are made possible by Fluke Calibration's exclusive quartz reference pressure transducer (Q-RPT) modules.

Q-RPTs measure pressure by measuring the change in the natural oscillating frequency of a quartz crystal with pressure induced stress. To be qualified for use in a Q-RPT module, each transducer is individually evaluated and characterized using automated primary pressure standards. Only transducers exhibiting required levels of linearity, repeatability and stability are selected. A proprietary compensation model, derived from more than 15 years experience with thousands of quartz pressure transducers, is applied to optimize the metrological characteristics needed in a transfer standard.

PPCH can be delivered with a low cost utility sensor for applications in which the high precision and stability of a Q-RPT are not required.

Q-RPTs available for PPCH
Q-RPT Designation SI Version Maximum Range
Absolute Gauge (MPa)
US Version Maximum
Range AbsoluteGauge (psi)
A200M1 200 30 000
A140M1 140 20 000
A100M1 100 15 000
A70M 70 10 000
A40M 40 6 000
A20M 20 3 000

1. Hi Q-RPT only, not available in Lo Q-RPT position.

Infinite Ranging™ and autoRange™

There is a lot more to covering a wide range of test devices with a single pressure controller than "% of reading" measurement uncertainty.

In addition to the necessary measurement uncertainty, PPCH offers the full pressure control and feature adaptability that are needed for true rangeability in test and calibration applications. Infinite ranging gives PPCH unprecedented versatility in adapting to a wide variety of devices to be tested. With the easy to use AutoRange function, a few simple key strokes or a single remote command string at the start of a test adapts every feature of the controller to optimize it for a specific range.

Open architecture

A PPCH controller can be configured with up to four Q-RPT modules. These can be internal or external to the PPCH controller. External Q-RPTs are in RPM4™ Reference Pressure Monitors. The RPM4's Q-RPTs then become part of the PPCH system and are managed by PPCH. External Q-RPTs must be disconnected or protected by valves when PPCH is used at pressure greater than the external Q-RPT range. Examples of possible PPCH system configurations include:

  • A PPCH with one or two built-in Q-RPTs to act as a stand alone, "one box" controller/calibrator package.
  • A PPCH with no internal Q-RPTs and an external Q-RPT to configure a system whose reference pressure measurement is remote from the controller. This configuration is ideal when it is advantageous for the reference to be removed from the system (e.g. for recalibration) while leaving the controller installed or to locate the reference measurement in closer proximity to the device or system under test.
  • A PPCH with no built-in Q-RPTs to act as a low cost automated pressure setting and controlling device (for example to automate pressure control in a PG7000™ piston gauge system).

Features, features, features

PPCH includes all the features you expect in today's state-of-the-art instruments and much more:

  • Pressure "ready/not ready" indicator with user adjustable criteria
  • Intelligent AutoZero™ function
  • 16 SI and US pressure units
  • Automatic fluid pressure head correction
  • On-board, programmable calibration sequences with DUT tolerance testing
  • Remote [ENTER] footswitch for hands free test execution
  • Valve drivers option for system design
  • Automated leak testing routines
  • RS-232 and IEEE- 488 communications
  • FLASH memory for simple and free embedded software upgrades

 

Specifications
PPCH General Specifications
Power Requirements 85 to 264 VAC, 50/60 Hz, 75 W max
Normal Operating Temperature Range 15 to 35 °C
Vibration Meets MIL-T-28800D
Weight (Typical) 50 kg approx (110 lb)
Dimensions 30 cm H x 52 cm W x 50 cm D
(12 in. x 20.5 in. x 20 in.) with enclosure
6U H rack mount version
Communications Ports RS232 (COM1, COM2), IEEE-488.2
Operating Modes Gauge, absolute
Pressure Ranges Atmosphere to 200 MPa (30,000 psi)
Operating Media Sebacate oil standard. Others optional (consult Fluke Calibration)
Internal Reservoir Volume 250 cc (external unlimited)
Drive Air Supply 70M - 500 kPa (75 psi), 300 l/m (10cfm)
140M - 500 kPa (75 psi), 450 l/m (15cfm)
100M - 700 kPa (100 psi), 300 l/m (10cfm)
200M - 700 kPa (100 psi), 450 l/m (15cfm)
Pressure Connections DRIVE AIR SUPPLY 1/8 in. NPT F
TEST DH500 (equivalent to AE F250C, HIP HF4)
Utility Sensor Precision/Resolution
± 0.10 % span / 0.001 % span
Drivers (8) 12V, 1 A max total output
CE Conformance Available, must be specified
Pressure Control
Control Modes Dynamic: Sets target pressure within hold limit and continuously adjusts pressure to remain at target value.

Static: Sets target pressure within hold limit and shuts off control, allowing pressure to stabilize naturally.

Monotonic: Sets pressure to target, then maintains very slow ramp in same direction as pressure increment.

Ramp: Sets and maintains user specified rate of change of pressure.

Piston Gauge: Control PPCH is controlled by PG7302™ to automate piston gauge pressure control.

Control Precision To ± 0.003 % of Q-RPT span
Control Volume 0 to 100 cc, 50 cc optimum
(operates in larger volumes but pressure stabilizing time increases)
Control Speed Slew rate (0 to full scale): 60 sec.

Dynamic mode typical time to Ready: 90 to 120 sec.
(Reduced by increasing hold limit or using monotonic control)

Lowest Controllable Pressure 1 MPa (150 psi)(lower in optimum conditions and with PG7302)
Measured and delivered pressure (Q-RPT)
Warm Up Time 30 minute temperature stabilization recommended from cold power up
Resolution To 1 ppm, user adjustable
Calibration A2LA accredited calibration report included
Q-RPTs Less than A200M A200M Less than A200M A200M
Precision1 ± 0.012 % of reading5 ±0.015 % of reading5
Predicted One Year Stability2 ± 0.005 % of reading ± 0.005 % of reading
Measurement Uncertainty Delivered 3 ± 0.013 % of reading5 ± 0.018 % of reading5
Pressure Uncertainty
(Dynamic Control Mode)4
± 0.016 % of reading5 ± 0.020 % of reading5
A20M 20 3 000

1. Combined linearity, hysteresis and repeatability.
2. Predicted one year stability limit (k=2) assuming regular use of AutoZero function. AutoZero occurs automatically when vented in     gauge mode, by comparison with a barometric reference in absolute mode. Absolute mode predicted one year stability without use     of AutoZ is ± (0.005 % Q-RPT span + 0.005 % of reading).
3. Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted one year     stability, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO "Guide to the     Expression of Uncertainty in Measurement."
4. Maximum deviation of the PPCH controlled pressure from the true value including measurement uncertainty and dynamic control     hold limt.
5. % of reading applies to 30 to 100 % of Q-RPT span. Under 30 % of Q-RPT span, uncertainty is a constant value obtained by     multiplying the % of reading value by 30 % of Q-RPT span.

 

Models and Accessories
Model Name Description
PPCH-nnnM PPCH Automated Pressure Controller/Calibrator for Hydraulic Pressure with no internal Q-RPT (utility sensor only)

To configure:

  • Determine maximum controlled pressure required. Select:
    • PPCH-200M for 200 MPa (30 000 psi)
    • PPCH-140M for 140 MPa (20 000 psi)
    • PPCH-100M for 100 MPa (15 000 psi)
    • PPCH-70M for 70 MPa (10 000 psi)
  • Configure an appropriate RPM4. Note that the maximum pressure of the PPCH defines the maximum pressure of the PPCH system.
PPCH-nnnM AnnnMc1/AnnnMC2 PPCH Automated Pressure Controller/Calibrator for Hydraulic Pressure with one or two internal Q-RPTs

To configure:

  • Determine maximum controlled pressure required. Select:
    • PPCH-200M for 200 MPa (30 000 psi)
    • PPCH-140M for 140 MPa (20 000 psi)
    • PPCH-100M for 100 MPa (15 000 psi)
    • PPCH-70M for 70 MPa (10 000 psi)
  • Select one or two Q-RPTs. Lo Q-RPT must be A70M or lower.
Accessory Description Description
A100M PPCH Quartz Reference Pressure Transducer (Q-RPT)
A140M PPCH Quartz Reference Pressure Transducer (Q-RPT)
A200M PPCH Quartz Reference Pressure Transducer (Q-RPT)
A40M PPCH Quartz Reference Pressure Transducer (Q-RPT)
A70M PPCH Quartz Reference Pressure Transducer (Q-RPT)
Footswitch Remote [ENTER] footswitch
RS232 Cable 9-pin, 2 m for PPC3 COM1 or PPC3-to-RPM4 connections
Reservoir External fluid reservoir
RPM4 Reference pressure monitor for external Q-RPTs
PPCH 08-SP Special fluid
PPC4 07 Special calibration
PPCH 06-01 Enclosure (for bench applications)
PPCH 05-01 CE mark
PPCH 04-02 SI units version
상호명 : (주)아이엠피티 | 주소 : 서울특별시 송파구 충민로 10(문정동 628) 가든파이브 툴관 3층 A-8호 | 사업자등록번호 : 220-81-22489 | Tel : 02.552.8582 | Fax : 02.553.0388 | E-mail : impt@impt.co.kr | 고객정보 수집동의 약관
COPYLIGHTⓒ 2021. IMPT CO.,LTD. ALL RIGHTS RESERVED.
Design by 위디지털