General |
Power requirements |
100 to 240 VAC 50/60 Hz, 70 W max consumption |
Operating temperature range |
15 to 40 °C (50 °F to 104 °F) |
Storage temperature range |
- 20 to 70 °C (-4 °F to 158 °F) |
Vibration |
Meets MIL-T-28800D |
Weight |
16.6 kg (36.5 lb) |
Dimensions |
PPC4 |
19 cm H x 35 cm W x 41 cm D (7.5 in. x 13.8 in. x 16.1 in.) |
PPC4-ui |
19 cm H x 35 cm W x 45 cm D (7.5 in. x 13.8 in. x 17.7 in.) |
Remote communication interfaces |
RS232 (COM1, COM2), USB (front panel) Optional: IEEE-488.2 or Ethernet and USB (Rear Panel) |
Languages Supported (Advanced UI and cockpit software) |
Chinese (simplified), Chinese (traditional), Czech, English, French, German, Italian, Japanese, Portuguese, Russian and Spanish |
Pressure ranges |
Vacuum to 14 MPa (2,000 psi) |
Operating medium |
Any clean, dry, non-corrosive gas |
Supply pressure |
Maximum desired set pressure + 70 kPa (10 psi) |
Exhaust pressure |
Atmosphere or vacuum for pressures under 5 psig (35 kPa gauge) |
Pressure connections |
TEST (+), TEST (-) |
1/8 in. NPT F |
SUPPLY |
1/8 in. NPT F |
EXHAUST |
3/8 in. NPT F |
ATM |
10-32 UNF |
Pressure limits |
Maximum working test pressure |
103% Hi Q-RPT maximum |
Maximum pressure on TEST port without damage |
115 % Hi Q-RPT maximum |
Utility sensor (if present) |
Resolution |
0.001% of span |
Precision |
0.1% of span |
Pressure Measurement |
Warm up time |
30 minute temperature stabilization recommended for best performance from cold power up. |
Resolution |
To 1 ppm, user adjustable |
Acceleration affect |
± 0.008 % /g maximum, worst axis. Allows operation at ± 20° from reference plane without significant effect. |
Predicted one year stability1 |
± 0.005% of reading (Gauge mode or Absolute mode with regular use of AutoZero) |
Precision2 |
Full Scale Standard Q-RPT Class ± 0.01 % of AutoRanged span4 Standard Q-RPT Class ± 0.008 % of reading5 Premium Q-RPT Class ± 0.005 % of reading7 |
Measurement Uncertainty3 |
Full Scale Standard Q-RPT Class ± 0.015 % of AutoRanged span4 Standard Q-RPT Class ± 0.01 % of reading6 Premium Q-RPT Class ± 0.008 % of reading8 |
1 Predicted Q-RPT measurement stability limit (k=2) over one year assuming regular use of AutoZero function. AutoZero occurs automatically in gauge mode whenever vented, by comparison with barometric reference in absolute mode. Absolute mode predicted one year stability without AutoZ is ± (0.005 % Q-RPT span + 0.005 % of reading).
2 Combined linearity, hysteresis, repeatability. Add + 1 Pa (0.00015 psi) in gauge mode with an Axxx (absolute) Q-RPT for the resolution and short term stability of the on-board barometer.
3 Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted one year stability limit, temperature effect and calibration uncertainty, combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in Measurement.
4 % of AutoRanged span, but with AutoRanged span no lower than 30 % of Q-RPT span (10 % with BG15K and G15K).
5 ±0.008 % of reading from 30 % to 100 % of Q-RPT span. Below 30 %; ±0.0024 % of Q-RPT span.
6 ±0.01 % of reading from 30 % to 100 % of Q-RPT span. Below 30 %; ±0.0030 % of Q-RPT span.
7 ±0.005 % of reading from 30 % to 100 % of any AutoRanged span between 30 % and 100 % of maximum Q-RPT span. Below 30 %; ±0.0015 % of AutoRanged span, or 0.0005 % of Q-RPT span, whichever is greater.
8 ±0.008 % of reading from 30 % to 100 % of any AutoRanged span between 30 % and 100 % of maximum Q-RPT span. Below 30 %; ±0.0024 % of AutoRanged span, or 0.0007 % of Q-RPT span, whichever is greater.
Note: Technical note 8050TN11 provides a detailed description of PPC4 Q-RPT uncertainties.
Pressure Control Modes and Ready Indication |
Static mode |
Sets pressure to target within hold limit and shuts off control in a closed test volume. Pressure is “ready when inside hold limit and stability test is met. |
Dynamic mode |
Sets pressure within hold limit and continuously adjusts pressure to remain at target value. Pressure is “ready when inside hold limit. |
Control parameters |
Hold limit, stability limit (default values can be adjusted by user) |
Pressure Control Performance |
Control precision |
± 4 ppm of Q-RPT span or ± 0.4 ppm of HI Q-RPT span, whichever is greater. |
Lowest controllable pressure in dynamic mode |
Gauge: Zero set by automated venting. Lowest point above or below zero limited only by Q-RPT resolution and control precision. Absolute, negative gauge: 2 kPa ±0.05% of Q-RPT span or 1 kPa (0.15 psia), whichever is greater |
Ultimate pressure (absolute, negative gauge) |
With low ultimate pressure option: 50 Pa (0.5 mbar, 0.007 psia) typical, depending on vacuum pump and connections Without low ultimate pressure option: 200 to 700 Pa (2 to 7 mbar, 0.03 to 0.1 psia) depending on vacuum pump and connections |
Typical pressure setting ready time (0.005 % hold limit, 50 cc test volume) |
15 to 35 s |
Slew time (ATM to FS with 50 cc test volume) |
30 s |
Typical test volume |
HI Q-RPT span: 0-2 MPa [0-300 psi]: 0-1,000 cc HI Q-RPT span: 3.5-14 MPa [500-2000 psi]: 0-500 cc |
Model Name |
Description |
|
PPC4-ui xxxxy/zzzzy |
PPC4 Controller/Calibrator with Advanced user interface (color display, full keypad, point-and-click knob)
To configure: Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class (y): Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Utility sensor Select a Lo Q-RPT (zzzz) if desired. Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Select options Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 For detailed information about configuration options, download the PPC4 product brochure. |
Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class (y): Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Utility sensor Select a Lo Q-RPT (zzzz) if desired. Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Select options Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 |
Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class (y): Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Utility sensor Select a Lo Q-RPT (zzzz) if desired. Full Scale Standard Q-RPT Standard Q-RPT Premium Q-RPT Select options Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 |
|
|
PPC4 xxxxy/zzzzy |
PPC4 Controller/Calibrator with Basic user interface (2 x 20 character, 10 key)
To configure: Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class Select a Lo Q-RPT (zzzz) if desired. Select options: Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 For detailed information about configuration options, download the PPC4 product brochure. |
Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class Select a Lo Q-RPT (zzzz) if desired. Select options: Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 |
Define the maximum controlled pressure required (up to 10 MPa) Select the Q-RPT or utility sensor (xxxx) for the maximum pressure identified and choose its class Select a Lo Q-RPT (zzzz) if desired. Select options: Units - SI or US nominal ranges CE compliance Remote interface - add IEEE-488.2 |
|
|
Accessory Description |
Description |
A1.4M |
PPC4 Quartz Reference Pressure Transducer (Q-RPT) SI Version Maximum Range (kPa) Absolute: 1,400 Maximum Range (kPa) Gauge: 1,400US Version Maximum Range (psi) Absolute: 200 Maximum Range (psi) Gauge:200 |
A100K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 110 Maximum Range (kPa) Gauge: 10 US Version Maximum Range (psi) Absolute: 16 Maximum Range (psi) Gauge: 1.5 |
A10M |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 10,000 Maximum Range (kPa) Gauge: 10,000 US Version Maximum Range (psi) Absolute: 1,500 Maximum Range (psi) Gauge: 1,500 |
A160K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 160 Maximum Range (kPa) Gauge: 60 US Version Maximum Range (psi) Absolute: 23 Maximum Range (psi) Gauge: 8 |
A200K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 200 Maximum Range (kPa) Gauge: 100 US Version Maximum Range (psi) Absolute: 30 Maximum Range (psi) Gauge: 150 |
A2M |
PPC4 Quartz Reference Pressure Transducer (Q-RPT) SI Version Maximum Range (kPa) Absolute: 2,000 Maximum Range (kPa) Gauge: 2,000US Version Maximum Range (psi) Absolute: 300 Maximum Range (psi) Gauge: 300 |
A3.5M |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 3,500 Maximum Range (kPa) Gauge: 3,500 US Version Maximum Range (psi) Absolute: 500 Maximum Range (psi) Gauge: 500 |
A350K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 350 Maximum Range (kPa) Gauge: 250 US Version Maximum Range (psi) Absolute: 50 Maximum Range (psi) Gauge: 35 |
A700K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 700 Maximum Range (kPa) Gauge: 700 US Version Maximum Range (psi) Absolute: 100 Maximum Range (psi) Gauge: 100 |
A7M |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 7,000 Maximum Range (kPa) Gauge: 7,000 US Version Maximum Range (psi) Absolute: 1,000 Maximum Range (psi) Gauge: 1,000 |
BA100K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: 7 to 110 Maximum Range (kPa) Gauge: n/a US Version Maximum Range (psi) Absolute: 10.2 to 16 Maximum Range (psi) Gauge: n/a |
BG15K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: n/a Maximum Range (kPa) Gauge:15 US Version Maximum Range (psi) Absolute: n/a Maximum Range (psi) Gauge: 2.2 |
G100K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: n/a Maximum Range (kPa) Gauge: 100 US Version Maximum Range (psi) Absolute: n/a Maximum Range (psi) Gauge: 15 |
G15K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: n/a Maximum Range (kPa) Gauge:15 US Version Maximum Range (psi) Absolute: n/a Maximum Range (psi) Gauge: 2.2 |
G200K |
PPC4 Quartz Reference Pressure Transducer (Q-RPT)
SI Version Maximum Range (kPa) Absolute: n/a Maximum Range (kPa) Gauge: 200 US Version Maximum Range (psi) Absolute: n/a Maximum Range (psi) Gauge: 30 |
RS232 |
9 pin, 2m for PPC4E COM1 |
BAS-SITE |
BAS-SITE COMPASS® for Pressure Software Enhanced, Site License |
BAS-SNGL |
BAS-SNGL COMPASS® for Pressure Software Basic, Single User |
ENH-SNGL |
ENH-SNGL COMPASS® for Pressure Software Enhanced, Single User |
PK-PPC-BG-DVU |
Dual volume unit for use with BG15K Q-RPT |
Reference vacuum kit, 110V |
VA-PPC/MPC-REF, 110V Vacuum pump package for PPC EXHAUST port |
Reference vacuum kit, 220V |
VA-PPC/MPC-REF, 220V Vacuum pump package for PPC EXHAUST port (220V) |
RPM4 |
Reference pressure monitor for external Q-RPTs and/or absolute mode AutoZ. |
SPLT |
Self purging liquid trap for PPC4 TEST port |